Intercomparison of Gafchromic™ films, TL detectors and TL foils for the measurements of skindose in Interventional Radiology (CROSBI ID 205621)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Kopeć, Renata ; Novák, Leos ; Carinou, Elefteria ; Clairand, Isabelle ; Dabin, Jeremie ; Datz, Hanan ; De Angelis, Cinzia ; Farah, Jad ; Huet, C ; Knežević, Željka et al. Intercomparison of Gafchromic™ films, TL detectors and TL foils for the measurements of skindose in Interventional Radiology // Radiation measurements, 71 (2014), 282-286. doi: 10.1016/j.radmeas.2014.04.008

Podaci o odgovornosti

Kopeć, Renata ; Novák, Leos ; Carinou, Elefteria ; Clairand, Isabelle ; Dabin, Jeremie ; Datz, Hanan ; De Angelis, Cinzia ; Farah, Jad ; Huet, C ; Knežević, Željka ; Järvinen, Hannu ; Majer, Marija ; Malchair, Francoise ; Negri, Anna ; Haruz Waschitz, Sigalit ; Siiskonen, Teemu ; Szumska, Agnieszka ; Trianni, Annalisa ; and Vanhavere, Filip

engleski

Intercomparison of Gafchromic™ films, TL detectors and TL foils for the measurements of skindose in Interventional Radiology

Several passive solid state dosemeters, such as GafchromicTM films and thermoluminesscence (TL) detectors, are used to estimate and monitor patient skin doses in interventional radiology. To determine the suitability of XR-TypeR GafchromicTM films and of detectors based on TL materials: pellets, chips and foils to measure skin dose, an intercomparison exercise has been organized within European Dosimetry Radiation Group - Working Group 12 “European Medical ALARA Network” (EURADOS WG12). To test response detectors were exposed to X-ray beams of energies and qualities applied clinically. A blind test was also performed to investigate the accuracy of the dose estimate by detectors exposed to unknown doses. We found the response of films to be strongly dependent on beam quality and filtration (increasing by up to 80 % with respect to reference beam quality). The response of TL detectors was found to be less dependent on beam quality (less than 25% variation), with TL foils showing less than 10% variation with respect to reference beam quality. To accurately estimate patient skin doses in interventional radiology it is important to choose the quality of the calibration beam to be as close as possible to the quality of beams actually applied in clinical work.

radiochromic films; thermoluminescence detectors (TLD); TL foils; sSkin doses in interventional radiology

Rad je presentiran na skupu 17th Solid State Dosimetry Conference (SSD17).

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

71

2014.

282-286

objavljeno

1350-4487

10.1016/j.radmeas.2014.04.008

Povezanost rada

Fizika, Kemija

Poveznice
Indeksiranost