Application of the Optical System Aramis for Determining the Stability of External Fixator (CROSBI ID 594080)
Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Pamuković, Frane ; Surjak, Martin ; Kodvanj, Janoš
engleski
Application of the Optical System Aramis for Determining the Stability of External Fixator
External fixators are used for simple, rapid and stable fixation of bone fragments and often represent a definitive treatment. Nearly all open and closed fractures of the long bones can be successfully treated with an external fixator. Biomechanical testing of external fixator is conducted to determine the stability of bone fragments which is very important as it provides better outcome with less complications. This paper presents the analysis of the stability and rigidity of the newly developed fixator. Two different types of tests were conducted, axial compression test and bending test, and the results were compared with results obtained on the external fixator Orthofix. The displacements were measured and calculated by using the non-contact optical measuring system Aramis.
External Fixator; Stability; Optical System Aramis
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
18-19.
2012.
objavljeno
Podaci o matičnoj publikaciji
Proceedings of the 29st Danubia-Adria Symposium on Advances in Experimental Mechanics
Ognjanović, Milosav
Beograd: Mašinski fakultet Univerziteta u Beogradu
978-86-7083-762-1
Podaci o skupu
Danubia-Adria Symposium on Advances in Experimental Mechanics
poster
26.09.2012-29.09.2012
Beograd, Srbija