Exploiting thermochromic materials for the rapid quality assurance of physiotherapy ultrasound treatment heads (CROSBI ID 183508)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Butterworth, Ian ; Barrie, Jill ; Zeqiri, Bajram ; Žauhar, Gordana ; Parisot, Benjamin Exploiting thermochromic materials for the rapid quality assurance of physiotherapy ultrasound treatment heads // Ultrasound in medicine & biology, 38 (2012), 5; 767-776. doi: 10.1016/j.ultrasmedbio.2012.01.021

Podaci o odgovornosti

Butterworth, Ian ; Barrie, Jill ; Zeqiri, Bajram ; Žauhar, Gordana ; Parisot, Benjamin

engleski

Exploiting thermochromic materials for the rapid quality assurance of physiotherapy ultrasound treatment heads

Significant nonuniformities in the acoustic intensity distribution generated by physiotherapy ultrasound treatment heads are not uncommon, potentially leading to significant localised temperature rises and tissue damage. An acoustic absorber tile containing a thermochromic pigment has been developed to provide rapid quality assurance of physiotherapy ultrasound treatment heads by virtue of a thermochromic colour change, indicating the time-averaged intensity distributions generated by these devices. As a bench-top device, the use of the tile is designed to mimic the nature of the physiotherapeutic application, requiring minimal training. Two designs where thermochromic pigments are added to the various polymeric layers of the tile are presented. Testing has been conducted with two physiotherapy treatment heads of differing performance, one of them notably exhibiting a strong ‘‘hot-spot’’ in localised acoustic time-averaged intensity. Findings show good qualitative agreement with classical hydrophone scans. Techniques are explored for the correction of nonlinearities in the thermochromic relationship, to enhance the accuracy of quantitative assessment.

ultrasound physiotherapy; quality assurance; thermochromic pigment; acoustic absorber; intensity distribution