Potentiometric titration of micromolar levels of anionic surfactants in model effluents using a sensitive potentiometric sensor (CROSBI ID 181788)
Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Galović, Olivera ; Samardžić, Mirela ; Derežić, Dragan ; Madunić-Čačić, Dubravka ; Sak-Bosnar, Milan
engleski
Potentiometric titration of micromolar levels of anionic surfactants in model effluents using a sensitive potentiometric sensor
Anionic surfactants have been potentiometrically determined at the ppm level using a sensitive surfactant sensor based on 1, 3-didecyl-2-methylimidazolium-tetraphenylborate as a sensing ion-exchanger. Standard solutions of 3-didecyl-2-methylimidazolium chloride and cetylpyridinium chloride at concentrations of 5 x 10-5 M were used as cationic titrants. The sensor exhibited Nernstian slope toward dodecylsulfate and dodecylbenzene sulfonate with slope values of 59.4 ± 0.2 and 58.9 ± 0.4 mV/concentration decade, respectively, and detection limits of 2 x 10-7 M. The obtained recoveries of anionic surfactants in model effluents varied between 96.4 and 97.0 % with a detection limit of 0.02 mg/L.
surfactant sensor; PVC membrane; anionic surfactant; potentiometric titration; model effluent