Modification Of Kosters Primary Gauge Block Interferometer (CROSBI ID 576355)
Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Katić, Marko ; Šimunović, Vedran ; Mudronja, Vedran
engleski
Modification Of Kosters Primary Gauge Block Interferometer
By definition, meter is expressed as radiation of a laser in vacuum. This definition of meter has to be transferred to a physical artifact and gauge blocks are calibrated by means of interferometry for just that purpose. Primary interferometers are used at the highest accuracy levels to perform calibration of national gauge block standards. These devices are very expensive, given the fact that they are manufactured on order and used almost exclusively by national measurement institutes. Therefore, LFSB decided to retrofit and modify an existing interferometer, and preliminary measurements show substantial improvements in accuracy.
Length measurement; interferometry; optical measurement methods
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
45-49.
2011.
objavljeno
Podaci o matičnoj publikaciji
13th International Scientific Conference on Production Engineering CIM 2011 / Biograd, Hrvatska, 2011-06-16/18
Podaci o skupu
13th International scientific conference on production engineering–CIM2011
predavanje
16.06.2011-18.06.2011
Biograd na Moru, Hrvatska