Factors affecting bio-impedance at high current pulse stimulation (CROSBI ID 473772)
Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Paćelat, Elvis ; Magjarević, Ratko ; Išgum, Velimir
engleski
Factors affecting bio-impedance at high current pulse stimulation
In transcranial electrical stimulation, electrodes with small area are used in order to selectively stimulate the motor cortex. We investigated the influence of pulse current stimuli parameters on the characteristics of the bio-impedance measured with cup EEG and corkscrew electrodes at high current densities. The impedance measured with corkscrew electrodes decreased slightly with pulse current intensity remaining stable with changes of other parameters. The impedance measured with cup electrodes was much more dependent on stimulation parameters and generally showed an increase whenever the charge density was increased.
bio-impedance; high current stimulation; transcranial stimulation
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
171-172-x.
1999.
objavljeno
Podaci o matičnoj publikaciji
Proceedings of the 11th Nordic-Baltic Conference on Biomedical Engineering
Hinrikus, Hiie ; Meigas, Kaju ; Lass, Jaanus ; Riipulk, Jevgeni
Talin: Peter Peregrinus Ltd
Podaci o skupu
11th Nordic-Baltic Conference on Biomedical Engineering
predavanje
06.06.1999-10.06.1999
Talin, Estonija