Testing of Medical Electrical EquipmentUltrasonic Therapy Equipment (CROSBI ID 473614)
Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Kovačić, Dubravko ; Ćurković, Božidar ; Magjarević, Ratko ; Breyer, Branko
engleski
Testing of Medical Electrical EquipmentUltrasonic Therapy Equipment
We have tested the output intensity of 9 ultrasonic therapeutic devices according to the safety requirements stated in the relevant international and at the national standard. Our results show that nearly 40% of transducer heads are not operating properly and another 46% should be recalibrated. Through this example we would like to point out the need for regular preventive maintenance of all electromedical equipment by profes-sional engineering staff clinical engineers. Preventive maintenance is the first and irreplacable step in safety assurance.
ultrasonic theraphy; mainternance; safety;clinical engineering; output power