The improvement of accuracy in 3D motion reconstruction systems (CROSBI ID 473608)
Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Pribanić, Tomislav ; Cifrek, Mario ; Tonković, Stanko
engleski
The improvement of accuracy in 3D motion reconstruction systems
This paper deals with residual of over-determined system after the calculation of direct linear transformation (DLT) parameters, its possible reduction and impact on overall reconstruction ac-curacy. A method for improvement of reconstruction accuracy based on selection of calibration points is presented. It will be shown that taking only those calibration points that do not significantly increase the over-all residual of the over-determined system improve the accuracy and decreases the number of calibration points needed.
3D reconstruction accuracy; direct linear transformation; calibration points
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
832-833-x.
1999.
objavljeno
Podaci o matičnoj publikaciji
Proceedings of the European Medical & Biological Engineering Conference EMBEC '99
Rehak, Peter ; Hutten, Helmut
Beč: Peter Peregrinus Ltd.
Podaci o skupu
EMBEC '99 - European Medical and Biological Engineering Conference
predavanje
04.11.1999-07.11.1999
Beč, Austrija