Gas Chromatographic Determination of Inorganic Gases in High Purity Ethylene (CROSBI ID 148720)
Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Jadrijević, Vesna ; Deur-Šiftar, Đurđica
engleski
Gas Chromatographic Determination of Inorganic Gases in High Purity Ethylene
A simple and sensitive gas chromatographic method using the single columm technique withouth an extreme working temperature range for the separation od O2, N2, CO and CO2 was developed. The analysis was performed on micropacked columns filled with active charcoal with particle size distribution smaller than 20 micrometer, using a microthermal conductivity detector. The results obtained show that such a columm is capable of good separation of O2, N2 and CO at room temperature. To shorten the elution time of CO2 the temperature has to be elevated above ambient tempoerature by about 100 C. Employing for ethylene removal the subtraction method introduced in gas chromatography very early by Ray, the achieved separation of inorganic gases was usefully applied for the detemination of inorganic gases in high purity ethylene.
gas cheomatography ; inorganic gases ; high purity ethylene
Rad je kao predavanje prezentiran na skupu 3. jugoslovanski kongres za čiso in uporabno kemijo, održanom od 12.-17.96.1972 g., . Ljubljana, Jugoslavija ; objavljen u Knjizi sažetaka, str. 13. (S/09).
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano