Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Solution to the ion feedback problem in Hybrid Photon Detectors and Photo Multiplier Tubes (CROSBI ID 85628)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Ferenc, Daniel ; Hrupec, Dario ; Lorenz, Eckart Solution to the ion feedback problem in Hybrid Photon Detectors and Photo Multiplier Tubes // Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 427 (1999), 3; 518-523. doi: 10.1016/S0168-9002(98)01403-X

Podaci o odgovornosti

Ferenc, Daniel ; Hrupec, Dario ; Lorenz, Eckart

engleski

Solution to the ion feedback problem in Hybrid Photon Detectors and Photo Multiplier Tubes

A general solution to the positive ion feedback problem in Hybrid Photon Detectors (HPD), photo multipliers (PM) and other similar detectors was found in the insertion of a permanent electrostatic potential barrier which prevents drift of positive ions from the anode, or the first dynode, towards the photocathode. In this paper we present the method as applied to the Intevac HPD.

hybrid photon detectors; ion feedback problem

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

427 (3)

1999.

518-523

objavljeno

0168-9002

10.1016/S0168-9002(98)01403-X

Povezanost rada

Fizika

Poveznice
Indeksiranost