Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Comparison of doses calculated by Monte Carlo method and measured by LiF TLD in the buildup region for a 60Co photon beam (CROSBI ID 144255)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Budanec, Mirjana ; Knežević, Željka ; Bokulić, Tomislav ; Mrčela, Iva ; Vrtar, Mladen ; Vekić, Branko ; Kusić, Zvonko Comparison of doses calculated by Monte Carlo method and measured by LiF TLD in the buildup region for a 60Co photon beam // Applied radiation and isotopes, 66 (2008), 12; 1925-1929. doi: 10.1016/j.apradiso.2008.04.021

Podaci o odgovornosti

Budanec, Mirjana ; Knežević, Željka ; Bokulić, Tomislav ; Mrčela, Iva ; Vrtar, Mladen ; Vekić, Branko ; Kusić, Zvonko

engleski

Comparison of doses calculated by Monte Carlo method and measured by LiF TLD in the buildup region for a 60Co photon beam

This work studied the percent depth doses of 60Co photon beams in the buildup region of a plastic phantom by LiF TLD measurements and by Monte Carlo calculations. An agreement within ± 1.5% was found between PDDs measured by TLD and calculated by the Monte Carlo method with the TLD in a plastic phantom. The dose in the plastic phantom was scored in voxels with thickness scaled by physical and electron density. PDDs calculated by electron density scaling showed a better match with PDD(MC, TLD) ; the difference is within ± 1.5% in the buildup region for square and rectangular field sizes.

Dose buildup ; LiF TLD ; Monte Carlo simulation

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

66 (12)

2008.

1925-1929

objavljeno

0969-8043

10.1016/j.apradiso.2008.04.021

Povezanost rada

Fizika, Kemija, Kliničke medicinske znanosti

Poveznice
Indeksiranost