Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi !

Electromagnetic Interference in the Primary Electromagnetic Laboratory (CROSBI ID 137310)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Malarić, Roman ; Ilić, Damir ; Malarić, Krešimir Electromagnetic Interference in the Primary Electromagnetic Laboratory // Measurement science review, 8 (2008), 1; 18-21

Podaci o odgovornosti

Malarić, Roman ; Ilić, Damir ; Malarić, Krešimir

engleski

Electromagnetic Interference in the Primary Electromagnetic Laboratory

This paper describes the measurements of the electromagnetic fields performed in the Primary Electromagnetic Laboratory at the Faculty of Electrical Engineering and Computing. Because of the many instruments present in the laboratory and the vulnerability to the ambient electromagnetic fields, it was interesting to investigate if the level of exposure to microwave frequencies exceeds the international and Croatian standards. The measurements were performed in the frequency range from 75 MHz to 3 GHz. The highest fields came from the nearby base station, but still less than 0.3 V/m, which is rather low.

electromagnetic interference; precise measurements; selective radiometer

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

8 (1)

2008.

18-21

objavljeno

1335-8871

Povezanost rada

Elektrotehnika