Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Potentiometric Response of a Commercial (Ag2S + PbS)-Membrane Selective Electrode to Hg(II) in Aqueous Solution (CROSBI ID 136443)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Radić, Njegomir Potentiometric Response of a Commercial (Ag2S + PbS)-Membrane Selective Electrode to Hg(II) in Aqueous Solution // Mikrochimica acta (1966), 85 (1985), 3-4; 209-218. doi: 10.1007/BF01198293

Podaci o odgovornosti

Radić, Njegomir

engleski

Potentiometric Response of a Commercial (Ag2S + PbS)-Membrane Selective Electrode to Hg(II) in Aqueous Solution

The preparation and performance of an Hg(II)-ISE are described. The analytical behavior of this electrode, made by chemical treatment of a commercial silver sulphide type Pb(II)-ISE, is discussed in terms of potential vs. concentration curves, potential vs. pH curves, and selectivity. The response to fixed cHg(II) as the pH is increased in the alkaline region corresponds to decrease in the free Hg2+ concentration down to pHg 17. The estimated limit of detection of total Hg(II) concentration is 10-5 M.

mercury(II) ion-selective electrode ; lead(II) ion-selective electrode ; alkaline medium

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

85 (3-4)

1985.

209-218

objavljeno

0026-3672

0026-3672

10.1007/BF01198293

Povezanost rada

Kemija

Poveznice
Indeksiranost