Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi !

Features of Voltage Pulse Plethysmography (CROSBI ID 530499)

Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija

Melinščak, Martina ; Šantić, Ante Features of Voltage Pulse Plethysmography // Proceedings of the Fifth IASTED International Conference on Biomedical Engineering / Gardner, J.W. (ur.). 2007. str. 302-307.-x

Podaci o odgovornosti

Melinščak, Martina ; Šantić, Ante

engleski

Features of Voltage Pulse Plethysmography

Measurement of tissue volume changes by voltage pulse plethysmography is being studied. In voltage pulse plethysmography short voltage pulses activate the bio-tissue and the characteristics of the transient process are measured. There are two possible implementations: with grounded bio-tissue and with grounded current sampling resistor. Measurement was carried out for both cases and the obtained wave-shapes for blood pulsations in the arteries and for respiration are presented. From the characteristics of measured voltage the parameters of equivalent circuit for electrode-skin interface with three elements are calculated for dry and gel-wetted electrodes.

voltage pulse plethysmography; current pulse plethysmography; electrode-tissue interface; tissue resistance variation

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

302-307.-x.

2007.

objavljeno

Podaci o matičnoj publikaciji

Proceedings of the Fifth IASTED International Conference on Biomedical Engineering

Gardner, J.W.

Podaci o skupu

The Fifth IASTED International Conference on Biomedical Engineering

predavanje

14.02.2007-16.02.2007

Innsbruck, Austrija

Povezanost rada

Elektrotehnika

Poveznice