A pulse driven low-power capacitive force transducer (CROSBI ID 470087)

Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija

Šantić, Ante ; Lacković, Igor ; Bilas, Vedran A pulse driven low-power capacitive force transducer // Proc. of the 8th International IMEKO Conference on Measurement in Clinical Medicine, BMI98 / Magjarević, Ratko (ur.). Zagreb: Hrvatsko društvo za komunikacije, računarstvo, elektroniku, mjerenja I automatiku (KoREMA), 1998. str. 2.7-2.10-x

Podaci o odgovornosti

Šantić, Ante ; Lacković, Igor ; Bilas, Vedran

engleski

A pulse driven low-power capacitive force transducer

A low power, two layer capacitive transducers for measurement of vertical force during gait are developed. The foamy rubber is used as a dielectric in the capacitive sensors because of its good linearity and low hysteresis. Three such sensors are placed under the shoe sole on each leg. To reduce the power consumption constant current short pulses with low duty cycle (1:40) were applied instead of an A.C. bridge with continuous sine wave power supply. That resulted in reduced current consumption for approximately twelve times (for 6-channel unit from 14mA to 1.2mA). Also, instead of using two 9V batteries for symmetrical supply, asymmetrical supply with one 9V battery was used. During the pulse, a ramp voltage, as a consequence of the sensor� s capacitance charging with a constant current, rises depending on the applied force. Close to the end of the pulse, a sample, inversely proportional to the applied force, is taken and kept by a sample&hold circuit. Micro-power operational amplifiers with very low current consumption were used for signal conditioning, so the output voltage of the transducer is calibrated and proportional to the applied force.

gait; vertical force; energy saving; capacitive sensor; current pulses; low-duty cycle

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

2.7-2.10-x.

1998.

objavljeno

Podaci o matičnoj publikaciji

Proc. of the 8th International IMEKO Conference on Measurement in Clinical Medicine, BMI98

Magjarević, Ratko

Zagreb: Hrvatsko društvo za komunikacije, računarstvo, elektroniku, mjerenja I automatiku (KoREMA)

Podaci o skupu

8th International IMEKO Conference on Measurement in Clinical Medicine, BMI'98

predavanje

16.09.1998-19.09.1998

Dubrovnik, Hrvatska