Energy dependence of new thermoluminescent detectors in terms of Hp(10) values (CROSBI ID 125564)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Miljanić, Saveta ; Knežević, Željka ; Štuhec, Matjaž ; Ranogajec-Komor, Maria ; Krpan, Katarina ; Vekić, Branko Energy dependence of new thermoluminescent detectors in terms of Hp(10) values // Radiation protection dosimetry, 106 (2003), 3; 253-256

Podaci o odgovornosti

Miljanić, Saveta ; Knežević, Željka ; Štuhec, Matjaž ; Ranogajec-Komor, Maria ; Krpan, Katarina ; Vekić, Branko

engleski

Energy dependence of new thermoluminescent detectors in terms of Hp(10) values

The aim of this work was to determine energy dependence characteristics in terms of values of the personal dose equivalent, HP(10). The following types of thermoluminescent detectors (TLDs) were investigated: (a) two new types based on lithium borate, Li_2B_4O_7:Cu, In and Li_2B_4O_7:Cu, In, Ag ; (b) two types based on the highly sensitive material LiF:Mg, Cu, P, TLD-700H and GR 200A ; (c) two well-known types of LiF:Mg, Ti detector, TLD-100 and TLD-700 and (d) highly sensitive Al:2O_3:C detectors. TLDs previously calibrated with Cs-137 gamma rays, were simultaneously irradiated with X ray beams in the range of mean energies between 33 and 116 keV. The irradiations were performed with detectors in polymethyl methacrylate (PMMA) holders placed on a 30 cm X 30 cm X 15 cm water phantom with PMMA walls (ISO phantom). Measured energy responses were compared with calculated data for Hp(10) values. The results confirmed the satisfactory tissue equivalent characteristics of all investigated TLDs except Al_2O_3:C, which (due to its large energy dependence) is suitable for personal dosimetry only with an appropriate filter.

energy dependence ; thermoluminescent detectors ; Hp(10)