Assessment of a new human liquid control material for use in quality control. Biochimica clinica 1988 ; 12: 1359-1364. (CROSBI ID 123139)
Prilog u časopisu | izvorni znanstveni rad
Podaci o odgovornosti
Topić, Elizabeta ; Mesić, Ranko
engleski
Assessment of a new human liquid control material for use in quality control. Biochimica clinica 1988 ; 12: 1359-1364.
An ethilene glycol-based human control serum was prepared and evaluated as a quality control material for cholesterol, triglyceride, uric acid, urea and creatinine determinations. The aliquots were stored at at various temperatures: room temperature, +4º ; ; C and -20º ; ; C. A good precision for a particular analysis was indicated by the results obtained on the intra-assay reproducibility in each of the series, also pointing to homogeneity of the material. The quality control material thus prepared is relatively inexpensive and appropriate for daily quality control of the cholesterol, triglyceride, uric acid, urea and creatinine determinations over a certain period of time.
quality control; cholestreol; triglyceride; uric acid; urea; creatinine; human control serum