Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Redox kinetics in cathodic stripping square-wave voltammetry (CROSBI ID 77748)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Lovrić, Milivoj ; Mlakar, Marina Redox kinetics in cathodic stripping square-wave voltammetry // Electroanalysis, 7 (1995), 12; 1121-1125. doi: 10.1002/elan.1140071205

Podaci o odgovornosti

Lovrić, Milivoj ; Mlakar, Marina

engleski

Redox kinetics in cathodic stripping square-wave voltammetry

Several redox systems with reactants which are immobilized at the electrode surface (EuSal complex, SeIV and SeIV in the presence of CuII, cysteine and p-nitrophenol) were studied by square-wave voltammetry to examine whether their redox kinetics can be measured by using the phenomenon of the ‘quasireversible maximum’. This is explained in detail. The following standard rate constants are obtained: EuSal complex ks = 720 ± 31 s−1, SeIV/CuII, log [ks/(cms−1)] = 0.94 ± 0.9, cysteine, log [ks/(cms−1)] = 0.6 ± 0.9 and p-nitrophenol, ks > 1200 ± 50 s−1.

reox kinetics ; square-wave voltammetry

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

7 (12)

1995.

1121-1125

objavljeno

1040-0397

1521-4109

10.1002/elan.1140071205