Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Improved Frequency Measurement by means of DMM and verification of its specifications (CROSBI ID 121021)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Hlupić, Nikica ; Butorac, Josip ; Krešić, Mario Improved Frequency Measurement by means of DMM and verification of its specifications // IEEE transactions on instrumentation and measurement, 54 (2005), 5; 1957-1963-x

Podaci o odgovornosti

Hlupić, Nikica ; Butorac, Josip ; Krešić, Mario

engleski

Improved Frequency Measurement by means of DMM and verification of its specifications

A simple, but highly accurate sampling method for the measurement of frequency by means of DMM or DVM is explained both theoretically and from the applicative point of view. The mathemathical background is given in short form, and comparison with other solutions is provided. High accuracy and behaviour regarding random errors are foreseen by computer simulation and then proved in real laboratory measurements. Finally, verification of factory time base specifications of sophisticated equipment is shown taking HP 3458A DMM as an example device under test.

measurement of frequency; sampling; digital multimeter (voltmeter)

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

54 (5)

2005.

1957-1963-x

objavljeno

0018-9456

Povezanost rada

Elektrotehnika, Računarstvo, Matematika

Indeksiranost