Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

A simple calibration method for 106Ru-106Rh eye applicators (CROSBI ID 117233)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Kovačević, Nenad ; Vrtar, Mladen ; Vekić, Branko A simple calibration method for 106Ru-106Rh eye applicators // Radiotherapy and oncology, 74 (2005), 293-299-x

Podaci o odgovornosti

Kovačević, Nenad ; Vrtar, Mladen ; Vekić, Branko

engleski

A simple calibration method for 106Ru-106Rh eye applicators

Ru-Rh eye applicators are used for the radiotherapy of eye malignancies such as melanomas. We present a method of dosimetry of these beta particle emitting applicators. Method is based on a Plexiglas phantom (constructed for this purpose) containing spherical shells and very small, 1 x 1 1 mm3 thermoluminescent dosimeters (TLD) as dosimeters. We determined 3-D depth doses and interpolated depth dose functions. Surface dose rate inhomogeneities and the consequences were considered and discussed. A possible influence of phantom component of the emission on the results was analysed. The method has overal combined uncertainty +-6% which is comparable, and slightly better, than other recent dosimetric methods.

106Ru-106Rh ophthalmic applicators; Thermoluminescent dosimeters (TLD); Choroidal melanoma; Beta dosimetry

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

74

2005.

293-299-x

objavljeno

0167-8140

Povezanost rada

Kemija

Indeksiranost