The design of narrow-band water loaded ultrasound transmitters (CROSBI ID 510342)
Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Ivančević, Bojan ; Petošić, Antonio ; Jambrošić, Kristian
engleski
The design of narrow-band water loaded ultrasound transmitters
The performance of thin-disk transducers made of PZT-5A ceramic using acoustic matching and air-backing is described using simplified 1-D model and experimental results. Several ultrasound transducers have been designed for studing influence of high power ultrasound energy on tissue. PZT-5A ceramic has been used as active element with thickness aproximatelly λ /2, depending on working frequency (500 kHz, 1MHz, 2 MHz) and with three different diameter to thickness ratio (D/t) for obtaining different directivity patterns in radiated medium. Theoretical input impedance, conductance and bandwidth of transducer systems has been compared with experimental results. Theoretical results have been derivied using Mason's equivalent circuit.
ultrasound transmitter; 1-D model; piezoceramic paramteres