Impact of different evaluation modes for the detection of clotting times on APTT results (CROSBI ID 509866)

Prilog sa skupa u časopisu | sažetak izlaganja sa skupa | međunarodna recenzija

Miloš, Marija ; Coen, Desiree ; Zadro, Renata Impact of different evaluation modes for the detection of clotting times on APTT results // Clinical chemistry and laboratory medicine. 2003. str. S144-S144

Podaci o odgovornosti

Miloš, Marija ; Coen, Desiree ; Zadro, Renata

engleski

Impact of different evaluation modes for the detection of clotting times on APTT results

Modern photo-optical coagulometers collect optical data over the entire course of clot formation, by monitoring changes in light absorbance or transmittance. By using different evaluation modes, they offer possibility to measure clot time at different points of clot formation. The aim of our study was to find out if there was a difference between aPTT results (n=104) obtained with different evaluation modes (fixed absorbance, drifting-baseline and point of inflexion) on Dade Behring coagulation analyzers BCS and BCT, by using Actin FS as aPTT reagent. Discrepant aPTT results were observed on all analyzed samples, being always longer with point of inflexion (mean value: 39.4 s, range: 20.5-99.1s) as compared with results obtained by using drifting-baseline (mean value: 33.3 s, range: 20.0-78.5 s) or fixed absorbance (mean value: 31.2 s, range: 19.0-80.8 s). Additionally, incorrect results were identified by visual inspection of reaction curves in 2 different situations. Extremely short aPTT results were obtained in 4 samples with fixed absorbance. Secondly, despite normal reaction curves in 5 samples, no clot result was measured by drifting mode. In all these cases, another 2 evaluation modes gave correct results. Differences between results obtained with different evaluation modes indicate the need to establish not only reagent-coagulometer specific, but also evaluation mode specific reference intervals. According to our experience we suggest verification of the reaction curve of all suspected results and, if needed, repeated testing with another evaluation mode.

evaluation mode; aPTT

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

S144-S144.

2003.

nije evidentirano

objavljeno

Podaci o matičnoj publikaciji

Clinical chemistry and laboratory medicine

978-88-3233-128-8

1434-6621

Podaci o skupu

IFCC-FESCC European Congress of Clinical Chemistry and Laboratory Medicine

poster

01.06.2003-05.06.2003

Barcelona, Španjolska

Povezanost rada

Temeljne medicinske znanosti

Indeksiranost