A microprocessor controlled gas-flowmeter with autocalibration and nonlinearity compensation (CROSBI ID 30728)
Prilog u knjizi | izvorni znanstveni rad
Podaci o odgovornosti
Marić, Ivan ; Gamberger, Dragan ; Bogunović, Nikola
engleski
A microprocessor controlled gas-flowmeter with autocalibration and nonlinearity compensation
The paper presents a microprocessor controlled gas-flowmeter comprising the autocalibration and the compensation of nonlinearities of the voltage to frequency converters. Depending upon required precision the calibration in two or more points is performed and generally nonlinear characteristic of the voltage to frequency converter can be piecewise linearly approximated. Owing to autocalibration a high accuracy of the flow calculation can be achieved even with low performance converters. No external adjustment of the converters are necessary.
Autocalibration, Nonlinearity compensation, Voltage to frequency converter, Microprocessor, Measurement quantity, Measurement system, Reference value