Model of total skin electron treatment using the "six-dual-field" technique (CROSBI ID 108699)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Faj, Dario ; Vrtar, Mladen ; Krajina, Zdenko ; Jurković, Slaven ; Margaretić, Damir Model of total skin electron treatment using the "six-dual-field" technique // Collegium antropologicum, 27 (2003), 2; 713-721-x

Podaci o odgovornosti

Faj, Dario ; Vrtar, Mladen ; Krajina, Zdenko ; Jurković, Slaven ; Margaretić, Damir

engleski

Model of total skin electron treatment using the "six-dual-field" technique

Durin implementation of the total skin electron treatment, using six-dual-field technique, at radiotherapy department a large number of measuremnts are needed. To assess depth dose curve required by clinicians and dose uniformity over a whole treatment plane, combinations of different irradiation parameters are used (electron energy, beam angle, scatterers). Measurements for each combination must be performed. One possible way to reduce number of measuremnts is to Model the treatment using the Monte Carlo simulation of electron transport. We made a simplified multiple-source Monte Carlo model of electron beam and tested it by comparing calculations and experimental results. Calculated data differs less than 5 percent from measurements in the treatment plane. During the treatment patient can be approximated using cylinders with different diameters and orientations. We tried to model the depth dose variations in the total skin electron treatment not just around the body cross-section (simplified to cylinders of different daimeters), but also along the body to account for the variations in body curvature longitudinally. This effect comes down to the problem of modeling distribution in different cylinders, but varying the longitudinal orientation of those cylinders. We compared Monte Carlo calculations and film measurements of depth dose curves for two orientations of the cylindrical phantom, which were the simplest for experimental arrangement. Comparison of the results proved accuracy of the model and we used it to calculate depth dose curves for a number of other cylinder orientations.

Radiotherapy; Mycosis fungoides; Skin; Electron; Model

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

27 (2)

2003.

713-721-x

objavljeno

0350-6134

Povezanost rada

Fizika, Kliničke medicinske znanosti, Javno zdravstvo i zdravstvena zaštita