Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Isolation of iron and strontium from liquid samples and determination of 55Fe and 89, 90Sr in liquid radioactive waste. (CROSBI ID 108004)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Grahek, Željko ; Rožmarić Mačefat, Martina Isolation of iron and strontium from liquid samples and determination of 55Fe and 89, 90Sr in liquid radioactive waste. // Analytica chimica acta, 511 (2004), 2; 339-348-x

Podaci o odgovornosti

Grahek, Željko ; Rožmarić Mačefat, Martina

engleski

Isolation of iron and strontium from liquid samples and determination of 55Fe and 89, 90Sr in liquid radioactive waste.

This paper describes the method of isolating iron and strontium from liquid samples with a low concentration of ions that enables simple and rapid determination of 55Fe and 89, 90Sr. The method consists of binding (concentrating) Fe and Sr at the cation exchanger Amberlite IR-120, their elution from cation exchanger with 4M HCl or 8M HNO3, isolating Fe on the extraction chromatographic TRU column with 4M HCl or 8M HNO3, and isolating Sr on the Sr.Spec column with the mixture of 8M HNO3+2M HCl or 5M HNO3. After the isolation, 55Fe is determined by liquid scintillation counting with scintillation solution, while activity of 89, 90Sr is obtained by Cherenkov counting in 5M HNO3. It was shown that successive counting can be used for simultaneous determination of 89, 90Sr activity. The activity ratio of 89Sr/90Sr (up to 20:1) and vice versa does not impact the determination. 55Fe is also determined immediately after isolation. The measurements in alpha, beta mode can be used to verify any presence of alpha-emitter (americium) in the fraction of iron and to correct the result. The method was tested by determining 55Fe and 89, 90Sr in model samples and radioactive waste samples. The paper also shows that Fe and Zn can be bound to the TEVA and TRU resins from the solutions of HCl, HNO3 and mixture of HCl + HNO3. The binding strength depend on the type of resin and the concentration of the acid or the concentration of acids in the mixture. These resin and acids can be used for mutual separation of Fe and Zn and their separation from other elements.

iron; zinc; strontium; separation; radioactive isotope; extraction chromatography

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

511 (2)

2004.

339-348-x

objavljeno

0003-2670

Povezanost rada

Geologija, Kemija

Indeksiranost