Napredna pretraga

Pregled bibliografske jedinice broj: 15376

IBIC study of charge collection properties in Si(Li) detectors


Jakšić, Milko; Fazinić, Stjepko; Tadić, Tonči; Bogovac, Mladen; Bogdanović, Ivančica; Pastuović, Željko
IBIC study of charge collection properties in Si(Li) detectors // Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms, 138 (1998), 1327-1332 (međunarodna recenzija, članak, znanstveni)


Naslov
IBIC study of charge collection properties in Si(Li) detectors

Autori
Jakšić, Milko ; Fazinić, Stjepko ; Tadić, Tonči ; Bogovac, Mladen ; Bogdanović, Ivančica ; Pastuović, Željko

Izvornik
Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms (0168-583X) 138 (1998); 1327-1332

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Sažetak
The nuclear microprobe technique ion beam induced charge (IBIC) has been used to probe the charge collection efficiency across the front surface of Si(Li) X-ray detectors. IBIC was performed using 2.667 MeV ^6Li ions, avoiding the large penetration depth of the conventional proton-IBIC. The front surfaces of two Si(Li) detectors were scanned to image inhomogeneities of incomplete charge collection (ICC). The relationship between the X-ray peak shape of the tested detectors and the degree of incomplete charge collection is discussed.

Izvorni jezik
Engleski

Znanstvena područja
Fizika

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus