Validation of a Coil Impedance Model for Simultaneous Measurement of Electromagnetic Properties and Inner Diameter of a Conductive Tube (CROSBI ID 496571)

Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija

Vasić, Darko ; Bilas, Vedran ; Ambruš, Davorin Validation of a Coil Impedance Model for Simultaneous Measurement of Electromagnetic Properties and Inner Diameter of a Conductive Tube // From the Electrometer to the Networked Instruments: A Giant Step toward a Deeper Knowledge / Demidenko, Serge ; Ottoboni, Roberto ; Petri, Dario et al. (ur.). Piscataway (NJ): Institute of Electrical and Electronics Engineers (IEEE), 2004. str. 1406-1409-x

Podaci o odgovornosti

Vasić, Darko ; Bilas, Vedran ; Ambruš, Davorin

engleski

Validation of a Coil Impedance Model for Simultaneous Measurement of Electromagnetic Properties and Inner Diameter of a Conductive Tube

Remote-field eddy-current technique (RFEC) for nondestructive inspection of conductive tubes cannot differentiate inner from outer defects. Also, the measurement results are greatly affected by electromagnetic properties (electrical conductivity and magnetic permeability) of the tube. Therefore, additional, independent measurement of the electromagnetic properties as well as inner diameter is required. Commercial eddy-current systems successfully measure these parameters using multifrequency, multicoil configuration. In this paper, we investigate the feasibility of the electromagnetic properties and inner diameter measurement with only one coil excited at one frequency. We have developed an optimization algorithm for determination of the tube properties, based on a coil impedance model. Measurements have been performed on several tubes made of different materials. The consistency of the model with the measurement results has been confirmed for investigated frequency range (1 to 50kHz).

coil impedance model; magnetic permeability; electrical

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

1406-1409-x.

2004.

objavljeno

Podaci o matičnoj publikaciji

Demidenko, Serge ; Ottoboni, Roberto ; Petri, Dario ; Piuri, Vincenzo ; Chong Tad Weng, Dave

Piscataway (NJ): Institute of Electrical and Electronics Engineers (IEEE)

Podaci o skupu

IMTC 2004 Instrumentation and Measurement Technology Conference

predavanje

18.05.2004-20.05.2004

Como, Italija

Povezanost rada

Elektrotehnika