Determination of oxalate in urine, using an amperometric biosensor with oxalate oxidase immobilized on the surface of a chromium hexacyanoferrate-modified graphite electrode (CROSBI ID 103982)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Milardović, Stjepan ; Grabarić, Zorana ; Tkalčec, Mihael ; Rumenjak, Vlatko Determination of oxalate in urine, using an amperometric biosensor with oxalate oxidase immobilized on the surface of a chromium hexacyanoferrate-modified graphite electrode // Journal of AOAC International, 83 (2000), 1212-1217-x

Podaci o odgovornosti

Milardović, Stjepan ; Grabarić, Zorana ; Tkalčec, Mihael ; Rumenjak, Vlatko

engleski

Determination of oxalate in urine, using an amperometric biosensor with oxalate oxidase immobilized on the surface of a chromium hexacyanoferrate-modified graphite electrode

A novel enzymatic amperometric method is described for the determination of oxalic acid in urine. An amperometric biosensor was made by immobilizing oxalate oxidase on the surface of a chromium(III) hexacyanoferrate-modified graphite electrode by using a bovine serum albumin and glutaraldehyde cross-linking procedure. The enzyme biocatalyzes oxalate decomposition in the presence of oxygen into carbon dioxide and hydrogen peroxide, The oxalate concentration, which is proportional to the amount of hydrogen peroxide, was determined amperometricaily by measuring the current resulting in the reduction of hydrogen peroxide at a very low working potential (0.05 V versus the Hg \ Hg2Cl2 \ 3M KCl electrode), which minimized the influence of the possible interferences present in human urine. All experiments were performed with succinic buffer, pH 3.8, containing 0.1 M KCl and 5.4 mM ethylenediaminetetraacetic acid. In an aqueous solution of pure oxalic acid, the biosensor showed good linearity in a concentration range of 2.5-100 mu M without the use of a dialysis membrane. For untreated urine samples, a high correlation (R-2 = 0.9949) was Obtained between oxalate concentrations added to urine samples and oxalate recoveries calculated for determinations with the described oxalate biosensor.

Flow-injection analysis. Enzyme electrode. Hydrogen-peroxide. Ferrocene. Fructose. Samples. Sensor. Assay