DVM input and leakage resistance analysis in the DVM-based resistance measurement systems (CROSBI ID 103872)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Malarić, Roman ; Butorac, Josip ; Leniček, Ivan DVM input and leakage resistance analysis in the DVM-based resistance measurement systems // IEEE transactions on instrumentation and measurement, 51 (2002), 6; 1323-1327-x

Podaci o odgovornosti

Malarić, Roman ; Butorac, Josip ; Leniček, Ivan

engleski

DVM input and leakage resistance analysis in the DVM-based resistance measurement systems

The digitsal voltmeter (DVM) input and leakage resistances have been analysed in the DVM-based resistance measurement system using one or two DVMs. The results are given both theoretically and experimentally. Shunting errors due to DVM-input and leakage resistances were substantially reduced by using a proper measurment method. The method is being further investigated for measuring high-ohm standard resistors

Digital Voltmeters; Input resistance; leakage resistance; resistance measurement