Sepsityper® kit versus in-house method in rapid identification of bacteria from positive blood cultures by MALDI-TOF mass spectrometry (CROSBI ID 320677)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Perše, Gabrijela ; Samošćanec, Ivana ; Bošnjak, Zrinka ; Budimir, Ana ; Kuliš, Tomislav ; Mareković, Ivana Sepsityper® kit versus in-house method in rapid identification of bacteria from positive blood cultures by MALDI-TOF mass spectrometry // Life, 12 (2022), 11; 1744, 9. doi: 10.3390/life12111744

Podaci o odgovornosti

Perše, Gabrijela ; Samošćanec, Ivana ; Bošnjak, Zrinka ; Budimir, Ana ; Kuliš, Tomislav ; Mareković, Ivana

engleski

Sepsityper® kit versus in-house method in rapid identification of bacteria from positive blood cultures by MALDI-TOF mass spectrometry

In order to further accelerate pathogen identification from positive blood cultures (BC), various sample preparation protocols to identify bacteria with MALDI-TOF MS directly from positive BCs have been developed. We evaluated an in-house method in comparison to the Sepsityper® Kit (Bruker Daltonics, Bremen, Germany) as well as the benefit of an on-plate formic acid extraction step following positive signal by the BACTECTM FX system. Confirmation of identification was achieved using subcultured growing biomass used for MALDI-TOF MS analysis. A total of 113 monomicrobial positive BCs were analyzed. The rates of Gram-positive bacteria correctly identified to the genus level using in- house method and Sepsityper® Kit were 63.3% (38/60) and 81.7% (49/60), respectively (p = 0.025). Identification rates at species level for Gram- positive bacteria with in-house method and Sepsityper® kit were 30.0% (18/60) and 66.7% (40/60), respectively (p < ; 0.001). Identification rates of Gram-negative bacteria were similar with the in-house method and Sepsityper® Kit. Additional on-plate formic acid extraction demonstrated significant improvement in the identification rate of Gram- positive bacteria at both genus and species level for both in-house (p = 0.001, p < ; 0.001) and Sepsityper® Kit methods (p = 0.007, p < ; 0.001). Our in-house method is a candidate for laboratory routines with Sepsityper® Kit as a back-up solution when identification of Gram-positive bacteria is unsuccessful.

MALDI-TOF MS ; blood culture ; identification ; in-house method ; Sepsityper kit