Direct quantification of PAHs and nitro-PAHs in atmospheric PM by thermal desorption gas chromatography with electron ionization mass spectroscopic detection (CROSBI ID 319122)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Drventić, Ivana ; Šala, Martin ; Vidović, Kristijan ; Kroflič, Ana Direct quantification of PAHs and nitro-PAHs in atmospheric PM by thermal desorption gas chromatography with electron ionization mass spectroscopic detection // Talanta, 251 (2023), 123761, 8. doi: 10.1016/j.talanta.2022.123761

Podaci o odgovornosti

Drventić, Ivana ; Šala, Martin ; Vidović, Kristijan ; Kroflič, Ana

engleski

Direct quantification of PAHs and nitro-PAHs in atmospheric PM by thermal desorption gas chromatography with electron ionization mass spectroscopic detection

In this work, we developed and optimized a method for the analysis of PAHs and nitro-PAHs in atmospheric particulate matter (PM) samples by using thermal desorption gas chromatography coupled with electron ionization single quadrupole mass spectrometry (TD-GC-(EI)-MS). The method uses thermal desorption from a PM on a filter sample as means of sample introduction to a column and obviates the need for complex extraction procedures, which are time-consuming and require environmentally unfriendly solvents. Moreover, the possibility of systematic errors is minimized and a significantly smaller amount of sample is required compared to traditional techniques requiring a pre-extraction step (approx. 10-times). Thirteen PAHs and three nitro-PAHs were used during method development. Although Tenax cartridges are typically used to capture volatile pollutants from the air, we found that glass-wool liner is the most suitable trap for the examined analytes after desorption from a quartz filter. Among the various instrument parameters which were tested and optimized, TD desorption flow and hold time, and temperature of the cooled injection system (CIS) proved to be most critical. We also found out that the matrix effect is especially pronounced in the case of high PM loadings, which should be kept in mind when planning the analysis. After the optimization, standard reference materials (ERM-CZ100 and NIST 1648a) were used for partial method validation and finally, real PM10 and PM2.5 samples from two Slovenian cities were successfully analyzed.

Nitrated polycyclic aromatic hydrocarbons ; Particulate matter ; TD-GC-MS ; Air pollution monitoring ; Green analytical method

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

251

2023.

123761

8

objavljeno

0039-9140

1873-3573

10.1016/j.talanta.2022.123761