OPTIMIZATION OF LC-QqQ-MS/MS METHOD FOR QUANTIFICATION OF ANDROGRAPHOLIDES IN HUMAN PLASMA (CROSBI ID 727434)

Prilog sa skupa u zborniku | sažetak izlaganja sa skupa | međunarodna recenzija

Kovačić, Jelena ; Jeličić, Mario-Livio ; Amidžić Klarić, Daniela ; Mornar, Ana OPTIMIZATION OF LC-QqQ-MS/MS METHOD FOR QUANTIFICATION OF ANDROGRAPHOLIDES IN HUMAN PLASMA // SOLUTIONS IN CHEMISTRY / Kassal, Petar ; Meštrović, Ernest ; Namjesnik, Danijel et al. (ur.). Zagreb: Hrvatsko kemijsko društvo, 2022. str. 98-98

Podaci o odgovornosti

Kovačić, Jelena ; Jeličić, Mario-Livio ; Amidžić Klarić, Daniela ; Mornar, Ana

engleski

OPTIMIZATION OF LC-QqQ-MS/MS METHOD FOR QUANTIFICATION OF ANDROGRAPHOLIDES IN HUMAN PLASMA

Andrographolides are diterpenoids found in the plant Andrographis paniculata and are widely used in traditional medicine, exhibiting anti- inflammatory properties.1 In this research we developed and validated high-performance liquid chromatography coupled with electrospray ionization triple-quadrupole tandem mass spectrometry (LC-QqQ-MS/MS) method for quantification of andrographolides in human plasma. The chromatographic analysis of pharmacological active andrographolide (AND), neoandrographolide (NAND) and 14-deoxy11, 12- didehydroandrographolide (DAND) was performed on a 15 cm HSS Cyano column with a gradient mobile phase of ultra-pure water and acetonitrile with the addition of formic acid (0.1%) as a modifier at a flow rate of 0.8 mL/min in 21 min. The validation results showed acceptable linearity of the method (r>0.999) for all analytes within the calibration range 0.7–10 µg/mL. The acceptable accuracy of the method was indicated by recovery of 90–107 %. The limit of quantification was found to be between 2 and 70 ng/mL while the limit of detection was found between 0.6 and 20 ng/mL. Multiple reaction monitoring (MRM) mode was optimized to quantify data under monitoring precursor-product ion transitions of m/z 351.1 → 105.0, 351.5 → 91.0, 351.1 → 79.0 and 351.1 → 77.0 for AND, m/z 481.6  319.0, 481.6  301.2, 481.6  289.2, 481.6  95.1 for NAND and m/z 333.1  314.9, 333.1  297.2, 333.1  257.3, 333.1  91.0 for DAND (Figure 1). This advanced method is simple, selective, accurate and cost-effective. As a result, this assay can be easily transferred in clinical settings for pharmacokinetic studies and therapeutic drug monitoring programs. ACKNOWLEDGEMENTS This work has been supported in part by the Croatian Science Foundation under the project numbers [UIP-2017-05-3949, DOK-2021- 02-7922] and project FarmInova (KK.01.1.1.02.0021) funded by the European Regional Development Fund.

IBD ; LC-QqQ-MS/MS ; human plasma, andrographolides

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

98-98.

2022.

objavljeno

Podaci o matičnoj publikaciji

SOLUTIONS IN CHEMISTRY

Kassal, Petar ; Meštrović, Ernest ; Namjesnik, Danijel ; Ribić, Rosana ; Šekutor, Marina ; Tomišić, Vladislav, Usenik, Andrea

Zagreb: Hrvatsko kemijsko društvo

978-953-8334-05-4

Podaci o skupu

International Conference Solutions in Chemistry 2022

poster

08.11.2022-11.11.2022

Sveti Martin na Muri, Hrvatska

Povezanost rada

Farmacija