Dosimetric verification of intensity modulated radiotherapy (IMRT) treatment plans for prostate cancer patients (CROSBI ID 315625)
Prilog u časopisu | izvorni znanstveni rad | domaća recenzija
Podaci o odgovornosti
Mrčela, Iva ; Gregov, Marin ; Budanec, Mirjana ; Murgić, Jure ; Jakšić, Blanka ; Prpić, Marin ; Prgomet Sečan, Angela ; Fröbe, Ana
engleski
Dosimetric verification of intensity modulated radiotherapy (IMRT) treatment plans for prostate cancer patients
Intensity modulated radiotherapy (IMRT) has become widely used as a standard radiation therapy technique for a treatment of localized prostate cancer. Advancing from conformal radiotherapy (3D CRT) to a more complex IMRT technique brought the need for more thorough verification of the accuracy in the dose delivery. In this work we present the clinical workflow and the results of patient specific quality assurance (PSQA) procedures for 40 prostate cancer patients treated with step and shot IMRT, since its implementation in our routine clinical practice. PSQA procedures include dosimetric verification of each treatment plan with dedicated rotational phantom and high- resolution matrix detector system Octavius 4D (PTW Freiburg) which allows three-dimensional comparison of the calculated and delivered radiation dose distribution. Our results proved compliance with the universal tolerance limits recommended for those procedures (1), giving the confidence in safety of the treatment and a challenge for the adoption of more stringent constraints in the future.
prostate IMRT ; patient specific QA
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o izdanju
61 (Supplement 3)
2022.
21-27
objavljeno
0353-9466
1333-9451
10.20471/acc.2022.61.s3.3
Povezanost rada
Fizika, Temeljne medicinske znanosti