Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Verification of a 6-part differential haematology analyser Siemens Advia 2120i (CROSBI ID 314599)

Prilog u časopisu | stručni rad | međunarodna recenzija

Čičak, Helena ; Radišić Biljak, Vanja ; Šimundić, Ana-Maria Verification of a 6-part differential haematology analyser Siemens Advia 2120i // Biochemia medica, 32 (2022), 2; 020710, 7. doi: 10.11613/BM.2022.020710

Podaci o odgovornosti

Čičak, Helena ; Radišić Biljak, Vanja ; Šimundić, Ana-Maria

engleski

Verification of a 6-part differential haematology analyser Siemens Advia 2120i

Introduction: The aim of this study was to perform a comprehensive verification of a 6-part differential haematology analyser Siemens Advia 2120i (Erlangen, Germany), prior to its routine implementation. Materials and methods: Our verification protocol included: precision (within- and between-run), estimated bias (%) as measure of trueness, which was calculated from observed and manufacturers’ declared value, analytical measuring interval (AMI), carryover, confirmation of a limit of blank (LoB), determination of a limit of detection (LoD) and limit of quantitation (LoQ). The K2 ethylenediaminetetraacetic acid (EDTA) patients’ leftover samples were used for verification of analyser Advia 2021i. Acceptance criteria were based on manufacturer technical specifications (Siemens), 2016 state-of-the-art criteria (Vis and Huisman), and EFLM Biological Variation Database. Results: The within- and between-run precision were acceptable for all parameters and the lowest coefficients of variation were observed for mean corpuscular volume (MCV) (0.3% and 0.6%, respectively). Estimated bias was within the acceptance criteria for all parameters except for MCV (the estimated bias was 2.2% (acceptance criteria 2.0%). AMI was confirmed for all tested parameters (r > 0.99). The carryover estimates ranged from 0.1% for platelet count (Plt) to 0.6% for red blood cell count and were within the manufacturers’ specifications (≤ 1%). Manufacturers’ claims for LoB were confirmed for leukocytes, erythrocytes, haemoglobin, and platelets. The estimated LoD and LoQ were 0.05 x109 /L and 0.1 x109 /L for white blood cell count, while for Plt values were 2 x109 /L and 3 x109 /L, respectively. Conclusions: Analytical performance of the Siemens Advia 2120i meets predefined quality goals and is suitable for routine use in a clinical laboratory.

verification ; haematology analyser ; haematology ; blood cell count

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

32 (2)

2022.

020710

7

objavljeno

1330-0962

1846-7482

10.11613/BM.2022.020710

Povezanost rada

Kliničke medicinske znanosti

Poveznice
Indeksiranost