Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube? (CROSBI ID 313629)

Prilog u časopisu | izvorni znanstveni rad | domaća recenzija

Mandić, Sanja ; Mandić, Dario ; Lukić, Iva ; Rolić, Tara ; Horvat, Vesna ; Lukić, Maja ; Osvald, Silvija ; Šerić, Vatroslav Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube? // Biochemia medica, 30 (2022), 3; 446-456. doi: 10.11613/BM.2020.030704

Podaci o odgovornosti

Mandić, Sanja ; Mandić, Dario ; Lukić, Iva ; Rolić, Tara ; Horvat, Vesna ; Lukić, Maja ; Osvald, Silvija ; Šerić, Vatroslav

engleski

Test results comparison and sample stability study: is the BD Barricor tube a suitable replacement for the BD RST tube?

Introduction: The aim was to evaluate the BD Barricor tubes by comparison with the BD Rapid Serum Tubes (RST) through measuring 25 analytes and monitoring sample stability after 24 hours and 7 days. Materials and methods: Samples of 52 patients from different hospital departments were examined. Blood was collected in BD RST and BD Barricor tubes (Becton, Dickinson and Company, Franklin Lakes, USA). Analytes were measured by Beckman Coulter AU 480 (Beckman Coulter, Brea, USA), Dimension EXL (Siemens Healthcare Diagnostics, Newark, USA) and ARCHITECT i2000SR (Abbott Diagnostics, Lake Forest, USA). Between-tube comparison for each analyte was performed, along with testing analyte stability after storing samples at 4 °C. Results: BD Barricor tubes showed unacceptable bias compared to BD RST tubes for potassium (K) (- 4.5%) and total protein (4.4%). Analyte stability after 24 hours was acceptable in both tested tubes for most of analytes, except for glucose, aspartate aminotransferase (AST) and lactate dehydrogenase (LD) in BD Barricor and free triiodothyronine in BD RST sample tubes. Analyte stability after 7 days was unacceptable for sodium, K, calcium, creatine kinase isoenzyme MB, AST, LD and troponin I in both samples ; additionally for glucose, alkaline phosphatase and albumin in BD Barricor. Conclusion: All analytes, except K and total protein, can be measured interchangeably in BD RST and BD Barricor tubes, applying the same reference intervals. For most of the analytes, sample re-analysis can be performed in both tubes after 24 hours and 7 days, although BD RST tubes show better 7-day analytes stability over BD Barricor tubes.

BD Barricor tube, stability, serum, plasma, evaluation

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

30 (3)

2022.

446-456

objavljeno

1330-0962

1846-7482

10.11613/BM.2020.030704

Povezanost rada

Biotehnologija u biomedicini (prirodno područje, biomedicina i zdravstvo, biotehničko područje)

Poveznice
Indeksiranost