Optimization of the direct LSC method for determination of biogenic component in liquids by applying 14C (CROSBI ID 311117)
Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Krajcar Bronić, Ines ; Sironić, Andreja ; Barešić, Jadranka ; Lovrenčić Mikelić, Ivanka ; Borković, Damir
engleski
Optimization of the direct LSC method for determination of biogenic component in liquids by applying 14C
Determination of fraction of biogenic component in liquid fuels by a direct radiocarbon measurement in liquid scintillation counter (direct-LSC method) has been validated by participation in the international intercomparison exercise. All the results for samples with the standard quench parameter SQP(E) value above ≈ 650 were accepted. Highly quenched sample of used edible oil was diluted with the 14C-free petroleum ether to optimize the region of applicability. It was established that quantitative results were obtained for SQP(E) values above 700, while in the SQP(E) region between 700 and 600 only the qualitative results of fbio can be taken.
biogenic component ; liquid fuels ; radiocarbon method ; direct LSC ; optimization ; intercomparison