P73 - Assessment of different CT simulators used in radiotherapy treatment planning: regional multicentric study (CROSBI ID 309815)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Petrovic, B. ; Faj, D. ; Markovic, M. ; Paunovic, D ; Stankovic, J ; Krestic-Vesovic, J. ; Miskovic, I ; Cicarevic, K. ; Kasabasic, M., Bibic, J. ; Budanec, M. et al. P73 - Assessment of different CT simulators used in radiotherapy treatment planning: regional multicentric study // Physica medica, 92, Supplement (2021), S1-S282. doi: 10.1016/S1120-1797(22)00375-1

Podaci o odgovornosti

Petrovic, B. ; Faj, D. ; Markovic, M. ; Paunovic, D ; Stankovic, J ; Krestic-Vesovic, J. ; Miskovic, I ; Cicarevic, K. ; Kasabasic, M., Bibic, J. ; Budanec, M. ; Kralik, I. ; Galic S. ; Hrepic, D. ; Ibrisimovic, L. ; Davidovic, J. ; Kolarevic, G. ; Novakovic, T.

engleski

P73 - Assessment of different CT simulators used in radiotherapy treatment planning: regional multicentric study

Results: Field size differences were below 1 mm for all applicators and energies, except for the 2 cm applicator (the smallest one), for which the array error was 5 mm. The penumbrae were systematically larger (differences ranging from 1.6 mm to 4.1 mm) when measured using Starcheck. The symmetry values measured with the array agreed for applicators greater than 6 cm. Conclusions: The Starcheck array is suitable for the profile analysis of low-energy x-ray units with some limitations. The lower resolution of the matrix against films results in a smoothing of all profiles (figure). This effect increases when the applicator size decreases. It affects the values of the penumbra and the symmetry but not the field size except for the 2 cm applicator. Still, it can be used for constancy checks if properly referenced to films.

CT simulator, radiotherapy, conversion curve CT-to-ED

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

92, Supplement

2021.

S1-S282

objavljeno

1120-1797

1724-191X

10.1016/S1120-1797(22)00375-1

Povezanost rada

Fizika, Kliničke medicinske znanosti

Poveznice
Indeksiranost