The value of extended inflammatory parameters obtained on Sysmex XN-1000 haematology analyser as early laboratory indicators of COVID-19 (CROSBI ID 308141)
Prilog u časopisu | Pismo uredniku | međunarodna recenzija
Podaci o odgovornosti
Lapić, Ivana ; Brenčić, Tina ; Rogić, Dunja ; Pozaić, Petra ; Juričić, Gordana ; Honović, Lorena
engleski
The value of extended inflammatory parameters obtained on Sysmex XN-1000 haematology analyser as early laboratory indicators of COVID-19
The present aimed to evaluate the utility of extended inflammatory parameters as early laboratory indicators of COVID-19, that are readily available during each analysis of the complete blood count on Sysmex XN-1000 haematology analyzer. The obtained results revealed that among extended inflammatory parameters obtained through CBC analysis on Sysmex XN hematology analyser, parameters associated with NEUT activation are not specific for COVID-19 compared to respiratory or flu-like infections with similar presentation. Notably, the presence of AS-LYMP in COVID-19 patients was the most prominent finding, indicating that this parameter can, with acceptable diagnostic accuracy, distinguish COVID- 19 from non-COVID-19 patients at the early disease phase.
COVID-19 ; hematology ; inflammatory parameters ; Sysmex XN-1000
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o izdanju
82 (3)
2022.
257-259
objavljeno
0036-5513
1502-7686
10.1080/00365513.2022.2056859
Povezanost rada
Kliničke medicinske znanosti