Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi !

A Hygrometer Calibration Facility Based on Two Temperature Humidity Generator (CROSBI ID 466047)

Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija

Zvizdić, Davor ; Galaso, Ivan ; Šerfezi, Damir ; Kolobarić, Vladimir A Hygrometer Calibration Facility Based on Two Temperature Humidity Generator // Papers and Abstracts from the Third International Symposium on Humidity&Moisture - Volume 1 / Stephanie, Bell (ur.). Teddington: NPL - National Physical Laboratory, 1998. str. 119-126-x

Podaci o odgovornosti

Zvizdić, Davor ; Galaso, Ivan ; Šerfezi, Damir ; Kolobarić, Vladimir

engleski

A Hygrometer Calibration Facility Based on Two Temperature Humidity Generator

Abstract: To respond to the increasing demand for traceable calibration of hygrometers in Croatia, a hygrometer calibration facility has been developed. It is based on two temperature recirculating humidity generator intended for operation in the 4°C to 60°C dew point range. The design of the facility is a compromise one, having to accomodate dew-point hygrometers, relative humidity meters and even chart recorders without remote sensors, i.e. the need for different calibration routines and levels of accuracy. The facility can , therefore, operate in three conceptualy different modes: as stand alone humidity generator, as a slave system directly controlled by transfer standard hygrometer and as an environmental chamber for comparison calibrations. The whole system is integrated and automated via the custom developed, graphycally interfaced software platform. The main design features of the facility are described as well as considerations regarding behaviour and the accuracy of the system.

hygrometers; calibration

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

119-126-x.

1998.

objavljeno

Podaci o matičnoj publikaciji

Papers and Abstracts from the Third International Symposium on Humidity&Moisture - Volume 1

Stephanie, Bell

Teddington: NPL - National Physical Laboratory

Podaci o skupu

Third International Symposium on Humidity&Moisture

poster

06.04.1998-08.04.1998

Teddington, Ujedinjeno Kraljevstvo

Povezanost rada

Strojarstvo