Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi !

Chronic electrical stimulation of peripheral nerves via deep-red light transduced by an implanted organic photocapacitor (CROSBI ID 306263)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Silverå Ejneby, Malin ; Jakešová, Marie ; Ferrero, Jose J ; Migliaccio, Ludovico ; Sahalianov, Ihor ; Zhao, Zifang ; Berggren, Magnus ; Khodagholy, Dion ; Đerek, Vedran ; Gelinas, Jennifer N et al. Chronic electrical stimulation of peripheral nerves via deep-red light transduced by an implanted organic photocapacitor // Nature biomedical engineering, (2022), 6; 741-753. doi: 0.1038/s41551-021-00817-7

Podaci o odgovornosti

Silverå Ejneby, Malin ; Jakešová, Marie ; Ferrero, Jose J ; Migliaccio, Ludovico ; Sahalianov, Ihor ; Zhao, Zifang ; Berggren, Magnus ; Khodagholy, Dion ; Đerek, Vedran ; Gelinas, Jennifer N ; Głowacki, Eric Daniel

engleski

Chronic electrical stimulation of peripheral nerves via deep-red light transduced by an implanted organic photocapacitor

Implantable devices for the wireless modulation of neural tissue need to be designed for reliability, safety and reduced invasiveness. Here we report chronic electrical stimulation of the sciatic nerve in rats by an implanted organic electrolytic photocapacitor that transduces deep-red light into electrical signals. The photocapacitor relies on commercially available semiconducting non-toxic pigments and is integrated in a conformable 0.1- mm3 thin-film cuff. In freely moving rats, fixation of the cuff around the sciatic nerve, 10 mm below the surface of the skin, allowed stimulation (via 50–1, 000-μs pulses of deep-red light at wavelengths of 638 nm or 660 nm) of the nerve for over 100 days. The robustness, biocompatibility, low volume and high-performance characteristics of organic electrolytic photocapacitors may facilitate the wireless chronic stimulation of peripheral nerves.

bioelectronics, implant, oganic semiconductor, chronic, photovoltaic

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

(6)

2022.

741-753

objavljeno

2157-846X

0.1038/s41551-021-00817-7

Povezanost rada

Biotehnologija u biomedicini (prirodno područje, biomedicina i zdravstvo, biotehničko područje), Fizika

Poveznice
Indeksiranost