Cardiac Tamponade after Right Ventricular Perforation Caused by Screw-in Lead (CROSBI ID 299886)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Giunio, Lovel ; Boric, Teo ; Bulat, Cristijan ; Dragicevic, Dragan ; Lozo, Mislav Cardiac Tamponade after Right Ventricular Perforation Caused by Screw-in Lead // International journal of angiology, 25 (2016), 05; e177-e179. doi: 10.1055/s-0036-1580700

Podaci o odgovornosti

Giunio, Lovel ; Boric, Teo ; Bulat, Cristijan ; Dragicevic, Dragan ; Lozo, Mislav

engleski

Cardiac Tamponade after Right Ventricular Perforation Caused by Screw-in Lead

A 69-year-old man was admitted after syncope followed with chest pain and signs of cardiac tamponade. He had undergone permanent dual-chamber pacemaker implantation 3 weeks earlier. Transthoracic echocardiography (TTE) confirmed a pericardial effusion, and urgent pericardial drainage was performed. Right ventricular perforation caused by active-fixation (screw-in) lead was verified by multislice computed tomography. The lead was extracted under fluoroscopy and bedside TTE monitoring in the operating room with cardiothoracic surgery backup. In the same act, the new ventricular passive-fixation lead was implanted.

► cardiac tamponade ► right ventricular perforation ► pacemaker ► screw-in lead ► complications ► pericardial drainage ► multislice computed tomography