Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi !

Comparison of new versus reused Harmonic scalpel performance in laparoscopic appendectomy in patients with acute appendicitis—a randomized clinical trial (CROSBI ID 299489)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Mihanović, Jakov ; Šikić, Nataša Lisica ; Mrklić, Ivana ; Katušić, Zvonimir ; Karlo, Robert ; Jukić, Miro ; Jerončić, Ana ; Pogorelić, Zenon Comparison of new versus reused Harmonic scalpel performance in laparoscopic appendectomy in patients with acute appendicitis—a randomized clinical trial // Langenbecks archives of surgery, 406 (2020), 1; 153-162. doi: 10.1007/s00423-020-02039-y

Podaci o odgovornosti

Mihanović, Jakov ; Šikić, Nataša Lisica ; Mrklić, Ivana ; Katušić, Zvonimir ; Karlo, Robert ; Jukić, Miro ; Jerončić, Ana ; Pogorelić, Zenon

engleski

Comparison of new versus reused Harmonic scalpel performance in laparoscopic appendectomy in patients with acute appendicitis—a randomized clinical trial

Background The aim of this study was to compare lateral thermal damage of the appendix and clinical outcomes after laparoscopic appendectomy using new versus reused Harmonic scalpels. Methods A total of 100 consecutive patients with acute appendicitis who underwent laparoscopic appendectomy were enrolled in the two-center, randomized clinical trial. Using a computer random number generator, patients were allocated to new or reused group. Histopathological measurement of lateral thermal damage of the appendiceal base and mesoappendix, speed of transection of the appendiceal base, duration of surgery, subjective rating of device functionality, length of hospital stay, and complications were compared within groups. Results The median lateral thermal damage on the appendiceal base in the new group (N = 49) was 0.2 mm (IQR 0.1–0.2) and 0.1 mm (IQR 0.1–0.3) in the reused group (N = 51) (P = 0.644), while on the mesoappendix for both groups, thermal damage was 0.1 mm (IQR 0.1–0.2) (P = 0.418). The median time required for base transection in both groups was 8 s (IQR 7–9) (P = 0.776). The median duration of surgery was also comparable between the groups (22 min, IQR 20–30 vs 25 min, IQR 21–35 ; P = 0.233). Two postoperative complications in the new group and one in the reused group were recorded (4% vs 2% ; P = 0.536). Surgeons’ subjective assessment of the instrument did not reveal significant difference between the groups in all of the investigated categories. Conclusions The results of our study support the reuse of Harmonic scalpels especially in the settings where economic constraints might hamper access to minimally invasive surgery to a larger number of patients. The results obtained on laparoscopic appendectomy might not be reproducible to other more demanding surgical procedures. Trial registration ClinicalTrials.gov registry under identifier NCT04226482

new ; used ; Harmonic scalpel ; laparoscopic appendectomy ; acute appendicitis

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

406 (1)

2020.

153-162

objavljeno

1435-2443

10.1007/s00423-020-02039-y

Povezanost rada

Kliničke medicinske znanosti

Poveznice
Indeksiranost