Comparison of Wavelight Allegretto Eye-Q and Schwind Amaris 750S excimer laser in treatment of high astigmatism (CROSBI ID 299363)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Bohac, Maja ; Biscevic, Alma ; Koncarevic, Mateja ; Anticic, Marija ; Gabric, Nikica ; Patel, Sudi Comparison of Wavelight Allegretto Eye-Q and Schwind Amaris 750S excimer laser in treatment of high astigmatism // Graefe's archive for clinical and experimental ophthalmology, 252 (2014), 10; 1679-1686. doi: 10.1007/s00417-014-2776-2

Podaci o odgovornosti

Bohac, Maja ; Biscevic, Alma ; Koncarevic, Mateja ; Anticic, Marija ; Gabric, Nikica ; Patel, Sudi

engleski

Comparison of Wavelight Allegretto Eye-Q and Schwind Amaris 750S excimer laser in treatment of high astigmatism

Purpose To compare functional outcomes of Wavelight Allegretto Eye-Q 400Hz and Schwind Amaris 750S excimer laser for astigmatism between 2 and 7 diopters(D). Methods Prospective comparative non-randomized case series of 480 eyes assigned in two laser groups and further divided into myopic and mixed astigmatism subgroups. All treatments were centered on corneal vertex. One-year results were compared between the groups. Statistical analysis was performed using z-test. Results Both Allegretto and Amaris postoperative uncorrected distance visual acuity (UDVA) improved in comparison to preoperative corrected distance visual acuity (CDVA). The difference was significant in the Allegretto group for myopic astigmatism (p=0.017). There was no difference in postoperative UDVA between lasers. Average sphere decreased in all groups for both lasers (p<0.001) without difference in effectiveness of spherical correction between lasers for both groups. In Allegretto, average cylinder decreased from −3.30D to −0.55D in myopic astigmatism (p<0.001) and from −3.84D to −0.85D in mixed astigmatism (p<0.001). In Amaris average cylinder decreased from −3.21D to −0.43D in myopic astigmatism (p<0.001) and from −3.66D to −0.58D in mixed astigmatism (p<0.001). Amaris group had less residual astigmatism (myopic astigmatism p= 0.023, mixed astigmatism p<0.001). Mean spherical aberration shifted from positive to negative in mixed astigmatism for both lasers. Conclusion Both lasers are effective in terms of UDVA, CDVA, spherical correction, and preservation of high-order aberrations. However, Amaris was more effective in cylinder correction.

Astigmatism . LASIK . Excimer laser . Ablation profile . High-order aberrations