Influence of Centricity of Force-Proving Instrument on Calibration Results (CROSBI ID 704895)
Prilog sa skupa u zborniku | stručni rad | međunarodna recenzija
Podaci o odgovornosti
Ćorić, Danko ; Meštrović, Krešimir ; Rodinger, Tomislav ; Šolić, Sanja
engleski
Influence of Centricity of Force-Proving Instrument on Calibration Results
For a valid measurement of the transferable force transducer, its correct positioning on the reference standard is important. In the article, the measurement of the portable Z4A/100 kN force transducer was carried out at different distances from the axial axis of the load (0 mm, 20 mm, 40 mm, 60 mm). The measurement results were processed and analyzed, from which it is evident that the values of the characteristic errors increase by moving away from the axial axis. This primarily refers to the reproducibility error, which increases by approximately 0.04% at all measurement points, and to a lesser extent to the reversibility error, for which an increase of 0.02% was recorded at lower force values.
Calibration ; force-proving instrument ; repeatability ; reproducibility ; hysteresis
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
161-170.
2021.
objavljeno
Podaci o matičnoj publikaciji
Ćorić, Danko ; Šolić, Sanja ; Ivušić, Franjo
Zagreb: Hrvatsko društvo za materijale i tribologiju (HDMT)
2459-5608
nije evidentirano
Podaci o skupu
21th International Conference on Materials (MATRIB 2021)
poster
30.06.2021-02.07.2021
Vela Luka, Hrvatska