Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Investigating the Prospect of Copper-Histidine Complex Modified Gold Electrode for Histidine and Human Serum Albumin Recognition (CROSBI ID 296378)

Prilog u časopisu | izvorni znanstveni rad

Kereković, Irena ; Domjanić Drozdek, Sandra Investigating the Prospect of Copper-Histidine Complex Modified Gold Electrode for Histidine and Human Serum Albumin Recognition // Journal of applied health sciences, 6 (2020), 1; 33-40. doi: 10.24141/1/6/1/4

Podaci o odgovornosti

Kereković, Irena ; Domjanić Drozdek, Sandra

engleski

Investigating the Prospect of Copper-Histidine Complex Modified Gold Electrode for Histidine and Human Serum Albumin Recognition

Histidine monolayer was adsorbed on the surface of the gold electrode and further tailored with Cu2+ cation in aim to investigate adsorption of histidine (His) and human serum albumin (HSA). Formation of the layer on the surface of the gold electrode was confirmed with quartz crystal microbalance. Binding of the His and HSA onto the modified electrode was successfully done for a wide range of tested concentrations. Electrode response was linearly proportional to concentration of His and HSA with correlation coefficient R2 = 0.9895 and R2 = 0.9952 respectively.

gold electrode ; copper ; histidine ; human serum albumin ; quartz crystal microbalance

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

6 (1)

2020.

33-40

objavljeno

1849-8361

2459-5640

10.24141/1/6/1/4

Povezanost rada

Kemija

Poveznice