Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Recommendations for the use of active personal dosemeters (APDs) in interventional workplaces in hospitals (CROSBI ID 295935)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

(EURADOS WG12) O’Connor, Una ; Carinou, Elefteria ; Clairand, Isabelle ; Ciraj-Bjelac, Olivera ; De Montee, Fracesca ; Domienik-Andrzejewska, Joanna ; Ferrari, Paolo ; Ginjaume, Merce ; Hršak, Hrvoje ; Hupe, Oliver et al. Recommendations for the use of active personal dosemeters (APDs) in interventional workplaces in hospitals // Physica medica, 87 (2021), 131-135. doi: 10.1016/j.ejmp.2021.05.015

Podaci o odgovornosti

O’Connor, Una ; Carinou, Elefteria ; Clairand, Isabelle ; Ciraj-Bjelac, Olivera ; De Montee, Fracesca ; Domienik-Andrzejewska, Joanna ; Ferrari, Paolo ; Ginjaume, Merce ; Hršak, Hrvoje ; Hupe, Oliver ; Knežević, Željka ; Sans Merce, Marta ; Sarmenton, Sandra ; Siiskonen, Teemu ; Vanhavere, Filip

EURADOS WG12

engleski

Recommendations for the use of active personal dosemeters (APDs) in interventional workplaces in hospitals

Occupational radiation doses from interventional procedures have the potential to be relatively high. The requirement to optimise these doses encourages the use of electronic or active personal dosimeters (APDs) which are now increasingly used in hospitals. They are typically used in tandem with a routine passive dosimetry monitoring programme, with APDs used for real-time readings, for training purposes and when new imaging technology is introduced. However, there are limitations when using APDs. A survey in hospitals to identify issues related to the use of APDs was recently completed, along with an extensive series of APD tests by the EURADOS Working Group 12 on Dosimetry for Medical Imaging. The aim of this review paper is to summarise the state of the art regarding the use of APDs. We also used the results of our survey and our tests to develop a set of recommendations for the use of APDs in the clinical interventional radiology/cardiology settings, and draw attention to some of the current challenges.

Radiation ; Protection ; Dosememter ; Interventional

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

87

2021.

131-135