Upotreba metode objektnog rastera za testiranje upotrebe kratkih mini implantata za retenciju donje pokrovne proteze
Upotreba metode objektnog rastera za testiranje upotrebe kratkih mini implantata za retenciju donje pokrovne proteze // 6. Međunarodni kongres Stomatološkog fakulteta Sveučilišta u Zagrebu. U: Acta stomatologica Croatica vol.54 br.2 / Klarić Sever, Eva (ur.).
Zagreb: Stomatolški fakultet Sveučilišta u Zagrebu, Hrvatsko Stomatološko društvo - Hrvatskog liječničkog zbora, 2020. str. 217-218 (poster, međunarodna recenzija, sažetak, znanstveni)
CROSBI ID: 1110614 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Upotreba metode objektnog rastera za testiranje upotrebe kratkih
mini implantata za retenciju donje pokrovne proteze
(Digital image correlation analysis of mandibular overdenture
retained by four short mini dental implants)
Autori
Kovačić, Ines ; Peršić Kiršić, Sanja ; Kranjčić, Josip ; Čelebić, Asja
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni
Izvornik
6. Međunarodni kongres Stomatološkog fakulteta Sveučilišta u Zagrebu. U: Acta stomatologica Croatica vol.54 br.2
/ Klarić Sever, Eva - Zagreb : Stomatolški fakultet Sveučilišta u Zagrebu, Hrvatsko Stomatološko društvo - Hrvatskog liječničkog zbora, 2020, 217-218
Skup
6. međunarodni kongres Stomatološkog fakulteta Sveučilišta u Zagrebu = 6th International Congress of the School of Dental Medicine University of Zagreb
Mjesto i datum
Zagreb, Hrvatska, 28.02.2020. - 29.02.2020
Vrsta sudjelovanja
Poster
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
extremely atrophied mandible ; mandibular overdenture ; short one-piece mini-implants ; digital image correlation
Sažetak
Introduction: Insertion of short mini dental implants (short MDIs) could be the new treatment modality for cases of extreme mandibular ridge resorption. Aim was to test the behavior of dental implants being both, slim and short (short MDIs) for complete over-denture (OD) retention by digital image correlation (DIC).Materials and methods: Based on CBCT scans of patients with extremely atrophied man-dible (interforaminal height < 10 mm), six identical models of an average extremely re-sorbed mandible were virtually built and 3D printed. In each model, four MDIs (6, 8 or 10 mm long ; 2 or 2.5 mm wide) were inserted interforaminally and CoCr reinforced over-dentures (OD) were loaded using a testing machine with 150N loads applied on the den-ture molar region bilaterally, unilaterally and the on anterior region. The precise measure-ment of von Mises Strain and vertical displacement was conducted using the DIC method. Results: Von Mises Strain showed very low values on all models ranging from 0.46 % to 1.23 %, with the highest values observed at the zone of bone-denture contact and at sites closer to the force application. The DIC showed, in all loading stages, identical spreading direction with the maximum strain value within the bone-denture contact line. The same mean value of vertical displacement was measured for all three types of loading, amount-ing to 0.2 mm on all the models. It followed that the vertical displacement was the result of denture subsidence due to “o-ring” elasticity.Conclusion: DIC has demonstrated a low level of uniformly distributed strain that allows the use of short MDIs for the retention and stabilization of mandibular OD.
Izvorni jezik
Hrvatski, engleski
Znanstvena područja
Dentalna medicina
POVEZANOST RADA
Projekti:
IP-2014-09-1218 - Definiranje mogućnosti uporabe mini dentalnih implantata (MDI) i njihovi rezultati u in vitro i u kliničkim prospektivnim istraživanjima (Mini implantati/ MDI) (HRZZ - 2014-09) ( CroRIS)
Ustanove:
Stomatološki fakultet, Zagreb
Profili:
Ines Kovačić
(autor)
Asja Čelebić
(autor)
Josip Kranjčić
(autor)
Sanja Peršić Kiršić
(autor)
Citiraj ovu publikaciju:
Časopis indeksira:
- Web of Science Core Collection (WoSCC)
- Emerging Sources Citation Index (ESCI)
- Scopus
- MEDLINE