Study on the uncertainty of passive area dosimetry systems for environmental radiation monitoring in the framework of the EMPIR “Preparedness” project   (CROSBI ID 290280)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Iurlaro, Giorgia ; Baranowska, Zuzana ; Campani, L ; Ciraj Bjelac, Olivera ; Ferrari, Paolo ; Knežević, Željka ; Majer, Marija ; Mariotti, F ; Morelli, B ; Neumaier, Stefan et al. Study on the uncertainty of passive area dosimetry systems for environmental radiation monitoring in the framework of the EMPIR “Preparedness” project   // Radiation measurements, 142 (2021), 106543, 10. doi: 10.1016/j.radmeas.2021.106543

Podaci o odgovornosti

Iurlaro, Giorgia ; Baranowska, Zuzana ; Campani, L ; Ciraj Bjelac, Olivera ; Ferrari, Paolo ; Knežević, Željka ; Majer, Marija ; Mariotti, F ; Morelli, B ; Neumaier, Stefan ; Nodilo, Marijana ; Sperandio, L ; Vittoria, F A ; Wołoszczuk, K ; Živanović, Miloš

engleski

Study on the uncertainty of passive area dosimetry systems for environmental radiation monitoring in the framework of the EMPIR “Preparedness” project  

One of the objectives of the EMPIR project 16ENV04 “Preparedness” is the harmonization of methodologies for the measurement of doses with passive dosimetry systems for environmental radiation monitoring in the aftermath of a nuclear or radiological event. In such cases, measurements are often performed at low radiation dose rates, close to the detection limit of the passive systems. The parameters which may affect the dosimetric results of a passive dosimetry system are analyzed and four laboratories quantitatively evaluate the uncertainties of their passive dosimetry systems. Typical uncertainties of five dosimetric systems in four European countries are compared and the main sources of uncertainty are analyzed using the results of a questionnaire compiled for this specific purpose. To compute the characteristic limits of a passive dosimetry system according to standard ISO 11929, the study of the uncertainty of the system is the first step. In this work the uncertainty budget as well as the characteristic limits (decision thresholds and detection limits) are evaluated and the limitations and strengths of a complete analysis of all parameters are presented

passive dosimetry systems ; uncertainty budget ; decision threshold ; detection limit ; environmental radiation monitoring ; emergency preparedness

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

142

2021.

106543

10

objavljeno

1350-4487

10.1016/j.radmeas.2021.106543

Povezanost rada

Fizika, Interdisciplinarne prirodne znanosti, Kemija

Poveznice