Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Spatially and Temporally Resolved Detection of Arsenic in a Capillary Dielectric Barrier Discharge by Hydride Generation High-Resolved Optical Emission Spectrometry (CROSBI ID 289452)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Burhenn, Sebastian ; Kratzer, Jan ; Svoboda, Milan ; Klute, Felix David ; Michels, Antje ; Veža, Damir ; Franzke, Joachim Spatially and Temporally Resolved Detection of Arsenic in a Capillary Dielectric Barrier Discharge by Hydride Generation High-Resolved Optical Emission Spectrometry // Analytical chemistry, 90 (2018), 5; 3424-3429. doi: 10.1021/acs.analchem.7b05072

Podaci o odgovornosti

Burhenn, Sebastian ; Kratzer, Jan ; Svoboda, Milan ; Klute, Felix David ; Michels, Antje ; Veža, Damir ; Franzke, Joachim

engleski

Spatially and Temporally Resolved Detection of Arsenic in a Capillary Dielectric Barrier Discharge by Hydride Generation High-Resolved Optical Emission Spectrometry

A new method for arsenic detection by optical emission spectrometry (OES) is presented. Arsine (AsH3) is generated from liquid solutions by means of hydride generation (HG) and introduced into a capillary dielectric barrier discharge (DBD) where it is atomized and excited. A great challenge in OES is the reduction of the recorded background signal, because it negatively affects the limit of detection (LOD). In conventional DBD/OES methods, the signal intensity of the line of interest, in this case arsenic, is integrated over a long time scale. However, due to the pulsed character of the plasma, the plasma on- time is only a small fraction of the integration time. Therefore, a high amount of noise is added to the actual signal in each discharge cycle. To circumvent this, in the present study the emitted light from the DBD is collected by a fast gated iCCD camera, which is mounted on a modified monochromator. The experimental arrangement enables the recording of the emission signal of arsenic in the form of a monochromatic 2D-resolved picture. The temporal resolution of the iCCD camera in the nanosecond range provides the information at which point in time and how long arsenic is excited in the discharge. With use of this knowledge, it is possible to integrate only the arsenic emission by temporally isolating the signal from the background. With the presented method, the LOD for arsenic could be determined to 93 pg mL–1 with a calibration curve linear over 4 orders of magnitude. As a consequence, the developed experimental approach has a potential for both mechanistic studies of arsine atomization and excitation in DBD plasmas as well as routine applications, in which arsenic determination at ultratrace levels is required.

Arsine ; optical emission spectrometry (OES)

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

90 (5)

2018.

3424-3429

objavljeno

0003-2700

1520-6882

10.1021/acs.analchem.7b05072

Povezanost rada

Fizika

Poveznice
Indeksiranost