THE ANALYSIS OF THE ACCURACY AND EVALUATION MEASUREMENT UNCERTAINTY OF CMM USING BALL BAR (CROSBI ID 282648)
Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Štrbac, Branko ; Hadžistević, Miodrag ; Runje, Biserka ; Horvatić Novak, Amalija ; Matin, Ivan ; Jotić, Goran
engleski
THE ANALYSIS OF THE ACCURACY AND EVALUATION MEASUREMENT UNCERTAINTY OF CMM USING BALL BAR
In order to estimate the accuracy of coordinate measuring machines (CMM), the recommendation of the ISO 10360-2 standard is to use gauge blocks, but other reference (calibration) gauges can be used. Besides the accuracy estimation, in this way the metrological traceability of current measuring systems is maintained and measurement uncertainty is determined, but only for these measurement tasks. Due to the complex hardware, the accuracy of these measuring systems varies from the orientation and length of the calibrated gauges. This paper presents a ball bar that is used as a reference gauge and CMM that is used for analyzing the influence of number of points for defining the center of the balls. In order to test the significance of parameters, full design of experiment (DOE) factors were used. Also, the measurement uncertainty of CMM was evaluated using the ISO 15530-3 guidelines for comparing measurement errors that are considered, for this measuring task, and the associated measurement uncertainty.
error ; measurement uncertainty ; CMM ; ball bar
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano